Special features
− Accuracy class 0.0005
− Simultaneous measurement with
optional 2 or 6 channels
− Resolution of measured signals
up to the physical limit
> 1,000,000 d
− Individual linearization
− Powerful digital filters
− Background calibration
− Operation via touchscreen,
keyboard or computer
− Improved temperature and EMC
behavior
− LAN/USB as required
Digital precision
measuring devices
DMP41
Working method of the devices
DMP41−T2 (2 simultaneous
measurement channels)
DMP41−T6 (6 simultaneous
measurement channels)
DSP
1.2345 mV/V
Ethernet
USB host (2 x)
SYNC IN
RS232
USB device (1 x)
Temperature
(1-wire)
Digital IN/OUT
4 4
D
A
SYNC OUT
4
Converter
Channel n
Channel 1
DSPD
A DSPD
A DSPD
A
.
.
.
.
.
.
B3479-2.0 en
D
at
a
sh
ee
t
Specifications
Type DMP41−T2 DMP41−T6
Accuracy class 0.00051)
Number of amplifiers
Transducers that can be connected
2
2 SG full bridges
6
6 SG full bridges
Excitation voltage (−) Ub
Carrier frequency
Transducer resistance per amplifier
at Ub = 2.5V
at 5V
at Ub = 10V
Length of transducer cable
Measuring ranges
Digital filter (6th order)
Display resolution
Common-mode rejection
Input resistance
V
Hz
m
mV/V
Hz
digit
dB
M
2.5; 5; 10
225 100ppm
75 … 4,000
150 … 4,000
300 … 4,000
< 200
2.5; 5, 10
40...0.01 (15 steps)
> 1,000,000
> 120
> 100
Sampling rate, per amplifier 1/s 1 … 450
Taring range/Zeroing range
Linearization of transducer characteristic curve
Total display range
2...11 points
Non-linearity
rel. to full scale value
Temperature effect per 10 K in nominal (rated) temperature
range
on zero point (rel. to full scale value)
on sensitivity (rel. to actual value)
Short-term drift over 5 min, from 2 h after switch-on
Long-term drift over 24 h, from 2 h after switch-on
%
%
%
ppm
ppm
< 0.0005
< 0.0002
< 0.0005
max. 2, typ. 1
max. 5, typ. 2
Nominal (rated) temperature range
Operating temperature range
Storage temperature range
C
C
C
10...+40
10...+50
-10...+60
Operating voltage (mains voltage) V 85 … 264 (50...60 Hz)
Power consumption
Weight, (net weight)
W
kg
approx. 35
approx. 9
approx. 45
approx. 9,5
Dimensions (W x H x D) mm 458 x 171 x 367
Connections for
SD transducer in six-wire configuration 2 x D-SUB-15
2 x Amphenol
6 x D-SUB-15
6 x Amphenol
Temperature sensor (1-wire), max. 4 sensors RJ45
Digital inputs and outputs
Computer interface Ethernet
Computer interface USB
USB host interface
Computer interface serial (optional)
D-SUB-15
RJ45
USB device
2 x USB Host
Adapter D-SUB-9
Range of use Indoor
Altitude, max. m 2000
Protection class I
Overvoltage category II
Permitted degree of contamination 2
Rel. air humidity, max. 80 % at 31 oC, decreasing linearly to 50% at 40oC
Degree of protection IP 20 per DIN EN 60529
1) With supply voltage 10V; Measuring range 2.5 mV/V; Transducer resistance 350 ; Cable length <10m
B3479−2.0 enHBM 2
Specifications (continued)
Scope of delivery
Article Order No.
Network cable (Ethernet cable CAT5+), 1 piece 1−KAB239−2
Network cable IEC 320 C13, 1 piece
Please enter the land in which the cable is to be used in the order (various
versions: DE/CH/GB/IT/USA)
1−KAB274− . . .
Synchronization cable for synchronization between 2 DMP41 devices (can be
used for both DMP41−T2 and DMP41−T6)
1−KAB261−2
Operating manuals, 2 copies −
1 pc USB cable, 2.0 (DMP41 − PC-connection) 3−3301.0127
Synchronization cable for synchronization between a DMP40 (can be used for
both DMP40 and DMP40−S2) and DMP41 devices (can be used for both
DMP41−T2 and DMP41−T6)
1−KAB298−0.5
Synchronization cable for synchronization between a DMP41 device and a
MGCplus device
1−KAB299−1
USB-RS232 adapter (from the new USB interface of the DMP41 to the RS232
interface previously present on the DMP40)
1−KAB297
Cable RS232 1−KAB2114−3
Accessories (not included among the items supplied)
Order No.
Temperature sensor (1-wire); with open ends; 1 sensor per temperature
channel
available from
www.wiregate.de
RJ45 connector for tool-free fitting 1−RJ45−EMV
Application areas
� Calibration measurements within scope of quality management for compliance with
DIN-ISO 9000, particularly for comparison measurements of comparison standards / test piece.
� Measurements under adverse conditions, extreme interference signal suppression.
� Precision measurement with SG transducers.
B3479−2.0 en HBM3
B
34
79
−
2.
0
en
Specifications (continued)
Housing rears
Digital inputs/outputs
Grounding switch
SG transducer, full bridge;
D-Sub male connector, 15pin,
DA−15P
RJ45 1−Wire
temperature
sensor
USB-device (PC)
USB host (RS232 adapter, mouse,
USB-stick etc.), e.g. for DMP40 connection
Ethernet network
Synchronization of
multiple DMP41
Housing
ground
Mains
connection
DMP41−T6
SG transducer, full bridge; MS cable
plug, 7pin, MS3106A 16S−P
DMP41−T2
SG transducer, full bridge; D-Sub
male connector, 15pin, DA−15P
SG transducer, full bridge; MS cable
plug, 7pin, MS3106A 16S−P
Digital inputs/outputs
USB-device (PC)
Housing
ground Mains connection
Grounding switch
USB host (RS232 adapter, mouse,
USB-stick etc.), e.g. for DMP40 connection
Ethernet networkSynchronization of
multiple DMP41
RJ45 1−Wire
temperature
sensor
Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45 64293 Darmstadt Germany
Tel. +49 6151 803-0 Fax: +49 6151 803-9100
Email: info@hbm.com www.hbm.com
Hottinger Baldwin Messtechnik GmbH.
Subject to modifications. All product descriptions are for
general information only. They are not to be understood as a
guarantee of quality or durability.
measure and predict with confidence