Upozornění: Text přílohy byl získán strojově a nemusí přesně odpovídat originálu. Zejména u strojově nečitelných smluv, kde jsme použili OCR. originál smlouvy stáhnete odsud
Příloha č.1 - Technická specifikace předmětu smlouvy č.
Ruční analyzátor spektra Rohde&Schwarz FPH 5 kHz - 13 GHz a FPH-B24 interní předzesilovač 5 kHz - 20 GHz
R&S®SPECTRUM RIDER FPH
HANDHELD SPECTRUM year
ANALYZER
Specifications
Data Sheet
Version 15.00
Version 15.00, May 2022
CONTENTS
Definitions ....................................................................................................................................................................... 3
Specifications.................................................................................................................................................................. 4
Frequency ......................................................................................................................................................................................... 4
Sweep time ....................................................................................................................................................................................... 4
Bandwidths........................................................................................................................................................................................ 5
Level ................................................................................................................................................................................................. 5
Trigger functions................................................................................................................................................................................ 9
Inputs and outputs ............................................................................................................................................................................10
General data.....................................................................................................................................................................................10
R&S®FPH-B100 type N RF input connector for model .26 ................................................................................................................12
R&S®FPH-B29 frequency extension from 5 kHz down to 100 Hz ......................................................................................................12
R&S®FPH-K7 modulation analysis....................................................................................................................................................12
R&S®FPH-K19 channel power meter................................................................................................................................................14
R&S®FPH-K29 pulse measurements with power sensor...................................................................................................................15
R&S®FPH-K43 receiver mode and channel scan measurement application......................................................................................15
R&S®FPH-K105 EMF measurement application ...............................................................................................................................16
R&S®FPH-K57 advanced gated trigger measurement ......................................................................................................................16
R&S®HA-Z350 log-periodic OEM antenna ........................................................................................................................................16
R&S®FSH-Z14 directional power sensor ..........................................................................................................................................17
R&S®FSH-Z44 directional power sensor ..........................................................................................................................................19
Equivalence of specifications for different R&S®Spectrum Rider FPH part numbers..........................................................................21
Ordering information .................................................................................................................................................... 21
Options .............................................................................................................................................................................................21
Extras ...............................................................................................................................................................................................22
Antennas and antenna accessories ..................................................................................................................................................22
Supported R&S®NRP-Zxx power sensors ........................................................................................................................................24
Optical power sensors and accessories............................................................................................................................................24
Service options.................................................................................................................................................................................25
2 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022Y-axis
Definitions
General
Product data applies under the following conditions:
• Three hours storage at ambient temperature followed by 30 minutes warm-up operation
• Specified environmental conditions met
• Recommended calibration interval adhered to
• All internal automatic adjustments performed, if applicable
Specifications with limits
Represent warranted product performance by means of a range of values for the specified parameter. These specifications are
marked with limiting symbols such as <, ≤, >, ≥, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by
testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift
and aging, if applicable.
Specification limit
Actual values with measurement uncertainty and guard band
Specification limit
Measurement uncertainties Guard band X-axis
Non-traceable specifications with limits (n. trc.)
Represent product performance that is specified and tested as described under “Specifications with limits” above. However, product
performance in this case cannot be warranted due to the lack of measuring equipment traceable to national metrology standards. In
this case, measurements are referenced to standards used in the Rohde & Schwarz laboratories.
Specifications without limits
Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent
values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is
ensured by design.
Typical data (typ.)
Characterizes product performance by means of representative information for the given parameter. When marked with <, > or as a
range, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents the
mean value.
Nominal values (nom.)
Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to
typical data, a statistical evaluation does not take place and the parameter is not tested during production.
Measured values (meas.)
Characterize expected product performance by means of measurement results gained from individual samples.
Uncertainties
Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been
calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account
environmental conditions, aging, wear and tear.
Device settings and GUI parameters are designated with the format “parameter: value”.
Non-traceable specifications with limits, typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.
In line with the 3GPP/3GPP2 standard, chip rates are specified in million chips per second (Mcps), whereas bit rates and symbol rates
are specified in billion bit per second (Gbps), million bit per second (Mbps), thousand bit per second (kbps), million symbols per
second (Msps) or thousand symbols per second (ksps), and sample rates are specified in million samples per second (Msample/s).
Gbps, Mcps, Mbps, Msps, kbps, ksps and Msample/s are not SI units.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 3
Version 15.00, May 2022 model .02 5 kHz to 2 GHz
with R&S®FPH-B3 option installed 5 kHz to 3 GHz
Specifications with R&S®FPH-B3 and R&S®FPH-B4 5 kHz to 4 GHz
options installed
Frequency 5 kHz to 6 GHz
model .06 5 kHz to 8 GHz
Frequency range with R&S®FPH-B8 option installed 5 kHz to 13.6 GHz
5 kHz to 20 GHz
Frequency resolution models .13/.23 (with tracking generator) 5 kHz to 26.5 GHz
Reference frequency, internal with R&S®FPH-B20 option installed 5 kHz to 31 GHz
Aging per year 5 kHz to 44 GHz
Temperature drift models .26/.36 (with tracking generator) from 5 kHz down to 100 Hz
Achievable initial calibration accuracy with R&S®FPH-B31 option installed
Total reference uncertainty 1 Hz
models .44/.54 (with tracking generator)
Frequency readout models .06/.13/.26/.23/.36/.44/.54,
Marker resolution with R&S®FPH-B29 option installed 1
Uncertainty 0 °C to +50 °C 1 × 10–6
Number of sweep (trace) points 1 × 10–6
Marker tuning frequency step size 5 × 10–7
Frequency counter resolution (time since last adjustment × aging rate) +
Count uncertainty temperature drift + calibration accuracy
Frequency span
Span uncertainty selectable 1 Hz
Spectral purity SNR > 25 dB ±(marker frequency × reference
SSB phase noise uncertainty + 10 % × resolution bandwidth
frequency = 500 MHz + ½ (span / (sweep points – 1) + 1 Hz)
Sweep time models .02/.06/.13/.26 711
span / 710
Sweep time carrier offset = 30 kHz 0.1 Hz (low), 0.1 mHz (high)
Uncertainty carrier offset = 100 kHz ±(frequency × reference uncertainty +
carrier offset = 1 MHz ½ (last digit))
models .23/.36/.44/.54 0 Hz,
carrier offset = 30 kHz 10 Hz to 2/3/4/6/8/13.6/20/26.5/31/44 GHz
carrier offset = 100 kHz 1 % (nom.)
carrier offset = 1 MHz
< –88 dBc (1 Hz), –95 dBc (1 Hz) (typ.)
< –98 dBc (1 Hz), –105 dBc (1 Hz) (typ.)
< –118 dBc (1 Hz), –125 dBc (1 Hz) (typ.)
< –88 dBc (1 Hz), –94 dBc (1 Hz) (typ.)
< –90 dBc (1 Hz), –96 dBc (1 Hz) (typ.)
< –115 dBc (1 Hz), –120 dBc (1 Hz) (typ.)
span = 0 Hz 56.8 µs to 1000 s
10 Hz ≤ span ≤ 600 MHz 20 ms to 1000 s
span > 600 MHz (20 ms × span / 1600 MHz) to 1000 s
span = 0 Hz 1 % (nom.)
span ≥ 10 Hz 3 % (nom.)
1 For serial number ≥ 103100. Not applicable to R&S®Spectrum Rider FPH model .02.
4 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Bandwidths –3 dB bandwidths 1 Hz to 3 MHz in 1/3 sequence
1 Hz ≤ RBW ≤ 300 kHz < 5 % (nom.)
Resolution bandwidths 300 kHz < RBW ≤ 1 MHz < 10 % (nom.)
Range < 5 (nom.) (Gaussian type filters)
Bandwidth accuracy –3 dB bandwidths
1 Hz to 3 MHz in 1/3 sequence
Selectivity 60 dB:3 dB
Video filters
Range
Level
Display range displayed noise floor to +30 dBm
Maximum rated input level with RF attenuation ≥ 10 dB
DC voltage 50 V
CW RF power model .02 33 dBm (= 2 W)
models .06/.13/.26 27 dBm (= 0.5 W)
models .23/.36/.44/.54 2 25 dBm (= 316 mW)
Peak RF power (duration < 3 s) model .02 36 dBm (= 4 W)
models .06/.13/.26/.23/.36/.44/.54 30 dBm (= 1 W)
models .23/.36/.44/.54 2 28 dBm (= 631 mW)
Maximum rated input level with RF attenuation < 10 dB
DC voltage 50 V
CW RF power model .02 20 dBm (= 100 mW)
models .06/.13/.26/.23/.36/.44/.54 20 dBm (= 100 mW)
Peak RF power (duration < 3 s) model .02 23 dBm (= 200 mW)
models .06/.13/.26/.23/.36/.44/.54 23 dBm (= 200 mW)
Intermodulation
Third order intercept (TOI) intermodulation-free dynamic range, signal level –20 dBm (both),
RF attenuation = 0 dB, RF preamplifier = off
model .02
f = 1 GHz +7 dBm (meas.)
f = 2.4 GHz +10 dBm (meas.)
models .06/.13/.26
f = 1 GHz +7 dBm (meas.)
f = 4.5 GHz, 22 GHz +8 dBm (meas.)
f = 9.5 GHz, 26.5 GHz +10 dBm (meas.)
f = 12 GHz +9 dBm (meas.)
models .23/.36/.44/.54
f = 1 GHz +10 dBm (meas.)
f = 4.5 GHz, 9.5 GHz, 26.5 GHz, +11 dBm (meas.)
32 GHz, 40 GHz
f = 12 GHz +8 dBm (meas.)
f = 22 GHz +9 dBm (meas.)
Second harmonic intercept (SHI) RF attenuation = 0 dB, RF preamplifier = off, signal level = –40 dBm
model .02
fin = 20 MHz to 1.5 GHz –60 dBc (nom.)
fin = 1.5 GHz to 2 GHz –80 dBc (nom.)
models .06/.13/.26
fin = 20 MHz to 1.5 GHz –60 dBc (nom.)
fin = 1.5 GHz to 14 GHz –90 dBc (nom.)
fin = 14 GHz to 15.4 GHz –85 dBc (nom.)
models .23/.36/.44/.54
fin = 20 MHz to 1.5 GHz –60 dBc (nom.)
fin = 1.5 GHz to 22 GHz –90 dBc (nom.)
2 RF input maximum power level ratings derate to +20 dBm for frequencies below 50 MHz.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 5
Version 15.00, May 2022 0 dB RF attenuation, termination 50 Ω, RBW = 1 kHz, VBW = 10 Hz,
Displayed average noise level (DANL)
sample detector, logarithmic scaling, normalized to 1 Hz
model .02
preamplifier = off
1 MHz to 10 MHz < –135 dBm, –142 dBm (typ.)
10 MHz to 1 GHz < –142 dBm, –146 dBm (typ.)
1 GHz to 4 GHz < –140 dBm, –144 dBm (typ.)
preamplifier = on
1 MHz to 10 MHz < –150 dBm, –160 dBm (typ.)
10 MHz to 3 GHz < –158 dBm, –163 dBm (typ.)
3 GHz to 4 GHz < –156 dBm, –161 dBm (typ.)
models .06/.13/.26
preamplifier = off
1 MHz to 10 MHz < –122 dBm, –130 dBm (typ.)
10 MHz to 25 MHz < –130 dBm, –135 dBm (typ.)
25 MHz to 1 GHz < –140 dBm, –145 dBm (typ.)
1 GHz to 19 GHz < –135 dBm, –140 dBm (typ.)
19 GHz to 27 GHz < –130 dBm, –138 dBm (typ.)
27 GHz to 29 GHz < –125 dBm, –130 dBm (typ.)
29 GHz to 31 GHz < –120 dBm, –123 dBm (typ.)
preamplifier = on
1 MHz to 20 MHz < –147 dBm, –152 dBm (typ.)
20 MHz to 3 GHz < –158 dBm, –162 dBm (typ.)
3 GHz to 4.5 GHz < –155 dBm, –158 dBm (typ.)
4.5 GHz to 27 GHz < –150 dBm, –155 dBm (typ.)
27 GHz to 29 GHz < –140 dBm, –145 dBm (typ.)
29 GHz to 31 GHz < –130 dBm, –133 dBm (typ.)
models .23/.36/.44/.54
preamplifier = off
1 MHz to 10 MHz < –125 dBm, –130 dBm (typ.)
10 MHz to 25 MHz < –130 dBm, –135 dBm (typ.)
25 MHz to 2.7 GHz < –140 dBm, –145 dBm (typ.)
2.7 GHz to 8 GHz < –135 dBm, –140 dBm (typ.)
8 GHz to 29 GHz < –133 dBm, –138 dBm (typ.)
29 GHz to 38 GHz < –130 dBm, –135 dBm (typ.)
38 GHz to 44 GHz < –125 dBm, –130 dBm (typ.)
preamplifier = on
1 MHz to 20 MHz < –147 dBm, –152 dBm (typ.)
20 MHz to 3 GHz < –157 dBm, –162 dBm (typ.)
3 GHz to 4.2 GHz < –150 dBm, –155 dBm (typ.)
4.2 GHz to 8 GHz < –153 dBm, –158 dBm (typ.)
8 GHz to 27.5 GHz < –145 dBm, –150 dBm (typ.)
27.5 GHz to 38 GHz < –140 dBm, –145 dBm (typ.)
38 GHz to 44 GHz < –130 dBm, –135 dBm (typ.)
6 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Immunity to interference model .02 –70 dBc (nom.)
Image frequencies fin – 2 × 30.15 MHz –70 dBc (nom.)
fin – 2 × 830.15 MHz –60 dBc (nom.)
Intermediate frequencies f < 3 GHz, fin – 2 × 4042.65 MHz –60 dBc (nom.)
f ≥ 3 GHz, fin + 2 × 830.15 MHz
–70 dBc (nom.)
models .06/.13/.26 –70 dBc (nom.)
fin – 2 × 30.15 MHz –50 dBc (nom.)
fin – 2 × 830.15 MHz –50 dBc (nom.)
f < 4 GHz, fin + 2 × 5582.35 MHz –70 dBc (nom.)
f < 4 GHz, fin + 2 × 7230.15 MHz
8 GHz ≤ f < 20 GHz, –70 dBc (nom.)
fin + 2 × 4030.15 MHz
8 GHz ≤ f < 20 GHz, –70 dBc (nom.)
fin + 2 × 5582.35 MHz
8 GHz ≤ f < 20 GHz, –40 dBc (nom.)
fin + 2 × 7230.15 MHz
20 GHz ≤ f < 26.5GHz, –60 dBc (nom.)
fin – 2 × 4030.15 MHz
26.5 GHz ≤ f < 28.5 GHz, –70 dBc (nom.)
fin – 2 × 7230.15 MHz –70 dBc (nom.)
–70 dBc (nom.)
models .23/.36/.44/.54 –70 dBc (nom.)
fin – 2 × 30.15 MHz –70 dBc (nom.)
fin – 2 × 830.15 MHz
f < 4.2 GHz, fin + 2 × 5582.35 MHz –70 dBc (nom.)
f < 4.2 GHz, fin + 2 × 7230.15 MHz
8 GHz ≤ f < 28 GHz, –70 dBc (nom.)
fin + 2 × 4030.15 MHz
8 GHz ≤ f < 28 GHz, –70 dBc (nom.)
fin + 2 × 5582.35 MHz
8 GHz ≤ f < 28 GHz, –70 dBc (nom.)
fin + 2 × 7230.15 MHz
28 GHz ≤ f < 44 GHz, –60 dBc (nom.)
fin – 2 × 5582.35 MHz
28 GHz ≤ f < 44 GHz, –60 dBc (nom.)
fin – 2 × 7230.15 MHz –50 dBc (nom.)
–40 dBc (nom.)
model .02
30.15 MHz, 830.15 MHz, 4042.65 MHz –50 dBc (nom.)
–60 dBc (nom.)
models .06/.13/.26
30.15 MHz, 830.15 MHz, 4030.15 MHz
5582.35 MHz
7230.15 MHz
models .23/.36/.44/.54
30.15 MHz
830.15 MHz, 4030.15 MHz,
5582.35 MHz, 7230.15 MHz
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 7
Version 15.00, May 2022
Other interfering signals, model .02 –60 dBc (nom.)
signal level – RF attenuation < –30 dBm f ≤ 3 GHz, spurious at
fin – 2021.325 MHz –60 dBc (nom.)
Other interfering signals, related to local –60 dBc (nom.)
oscillators models .06/.13/.26 –60 dBc (nom.)
f < 4 GHz, spurious at –60 dBc (nom.)
Residual spurious response fin + 2791.175 MHz –60 dBc (nom.)
f < 4 GHz, spurious at –60 dBc (nom.)
fin + 3615.075 MHz –60 dBc (nom.)
4 GHz ≤ f < 8 GHz, spurious at –60 dBc (nom.)
fin – 415.075 MHz
8 GHz ≤ f < 20 GHz, spurious at –60 dBc (nom.)
fin + 2015.075 MHz –60 dBc (nom.)
8 GHz ≤ f < 20 GHz, spurious at –60 dBc (nom.)
fin + 2791.175 MHz –60 dBc (nom.)
8 GHz ≤ f < 20 GHz, spurious at –60 dBc (nom.)
fin + 3615.075 MHz –60 dBc (nom.)
20 GHz ≤ f < 26.5 GHz, spurious at –60 dBc (nom.)
fin – 2015.075 MHz –60 dBc (nom.)
26.5 GHz ≤ f < 28.5 GHz, spurious at
fin – 3615.075 MHz –60 dBc (nom.)
–60 dBc (nom.) except otherwise stated
models .23/.36/.44/.54 –35 dBc (nom.)
f < 4.2 GHz, spurious at –40 dBc (nom.)
fin + 2791.175 MHz
f < 4.2 GHz, spurious at –60 dBc (nom.) except otherwise stated
fin + 3615.075 MHz –50 dBc (nom.)
4.2 GHz ≤ f < 8 GHz, spurious at –50 dBc (nom.)
fin – 415.075 MHz –50 dBc (nom.)
8 GHz ≤ f < 28 GHz, spurious at –40 dBc (nom.)
fin + 2015.075 MHz –40 dBc (nom.)
8 GHz ≤ f < 28 GHz, spurious at –50 dBc (nom.)
fin + 2791.075 MHz –90 dBm (nom.)
8 GHz ≤ f < 28 GHz, spurious at
fin + 3615.075 MHz
28 GHz ≤ f < 44 GHz, spurious at
fin – 2791.075 MHz
28 GHz ≤ f < 44 GHz, spurious at
fin – 3615.075 MHz
f = receive frequency
model .02
∆f ≥ 300 kHz
models .06/.13/.26
∆f ≥ 300 kHz, ∆f ≤ 1600 MHz
∆f ≤ –422.5 MHz,
21440 MHz ≤ fin < 23400 MHz
∆f ≥ 1115 MHz,
23400 MHz ≤ fin < 24400 MHz
models .23/.36/.44/.54
∆f ≥ 300 kHz
∆f ≥ 431 MHz,
11500 MHz ≤ fin ≤ 12000 MHz
∆f ≤ –230 MHz,
14000 MHz ≤ fin ≤ 14300 MHz
∆f ≥ 268 MHz,
14700 MHz ≤ fin ≤ 15500 MHz
∆f ≤ –187 MHz,
16500 MHz ≤ fin ≤ 17300 MHz
∆f ≤ –937 MHz,
18900 MHz ≤ fin ≤ 22000 MHz
∆f ≤ –337 MHz,
29900 MHz ≤ fin ≤ 31000 MHz
input matched with 50 Ω,
without input signal, RBW ≤ 30 kHz,
f ≥ 3 MHz, RF attenuation = 0 dB
8 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Level display 1/2/3/5/10/20/30/50/100/120/150 dB,
Logarithmic level axis 10 divisions
Linear level axis 0 % to 100 %, 10 divisions
Number of traces 2
Trace detectors max. peak, min. peak, auto peak, sample,
Trace functions RMS
Setting range of reference level clear/write, max. hold, min. hold, average,
Units of level axis view
Level measurement uncertainty –130 dBm to +30 dBm
Absolute level uncertainty at 100 MHz dBm, dBmV, dBµV, V, W
Frequency response 3
temperature range from +20 °C to +30 °C < 0.3 dB
Attenuator uncertainty
Uncertainty of reference level setting temperature range from +20 °C to +30 °C
Display nonlinearity
Bandwidth switching uncertainty 5 kHz ≤ f < 10 MHz < 1.5 dB (nom.)
Total measurement uncertainty
10 MHz ≤ f < 8 GHz 4 < 1 dB
Trigger functions
8 GHz ≤ f < 20 GHz < 1.5 dB
Trigger
Trigger source 20 GHz ≤ f ≤ 44 GHz < 2 dB
External trigger level threshold
< 0.3 dB
Gated trigger
Mode available < 0.1 dB (nom.)
Gate delay
Gate length SNR > 16 dB, 0 dB to –50 dB, < 0.3 dB
logarithmic level display
reference: RBW = 10 kHz < 0.1 dB (nom.)
95 % confidence level, +20 °C to +30 °C,
SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto
10 MHz ≤ f < 8 GHz 0.74 dB
8 GHz ≤ f < 20 GHz 1.04 dB
20 GHz ≤ f ≤ 44 GHz 1.35 dB
low → high transition free run, video, external
high → low transition 2.4 V
maximum 0.7 V
3.0 V
frequency mode, channel mode,
spectrogram mode
1 µs to 100 s, min. resolution: 1 µs
(or 1 % of delay)
1 µs to 100 s, min. resolution: 1 µs
(or 1 % of gate length)
3 For specifications with R&S®FPH-B100 option installed, see section R&S®FPH-B100 N type RF input connector for model .26.
4 10 MHz to 50 MHz frequency response is < 0.5 dB after alignment with R&S®FPH-K35 option.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 9
Version 15.00, May 2022
Inputs and outputs
RF input
Impedance 50 Ω (nom.)
type N, female
Connector models .02/.06/.13/.23 PC 3.5 mm male
PC 2.92 mm male
models .26/.36
< 1.5 (nom.)
models .44/.54 < 2 (nom.)
VSWR 3 model .02 < 2 (nom.)
< 1.5 (nom.)
100 kHz ≤ f ≤ 1 GHz < 2 (nom.)
1 GHz < f ≤ 4 GHz < 1.5 (nom.)
< 1.9 (nom.)
models .06/.13/.26 < 2.2 (nom.)
0 dB to 40 dB, in 5 dB steps
100 kHz ≤ f ≤ 100 MHz
AM and FM
100 MHz ≤ f ≤ 1 GHz 3.5 mm mini jack
32 Ω (nom.)
1 GHz < f ≤ 31 GHz adjustable from 0 V to > 100 mV (RMS)
models .23/.36/.44/.54 100 kHz to model maximum frequency
(usable to 30 kHz)
100 kHz ≤ f < 4.2 GHz –10 dBm to –30 dBm in 1 dB steps (nom.)
0 dBm to –25 dBm in 1 dB steps (nom.)
4.2 GHz ≤ f < 22 GHz –10 dBm to –30 dBm in 1 dB steps (nom.)
0 dBm to –25 dBm in 1 dB steps (nom.)
22 GHz ≤ f < 44 GHz –10 dBm to –30 dBm in 1 dB steps (nom.)
< 2.0 (nom.)
Input attenuator RF input only < 1.5 (nom.)
< 1.7 (nom.)
AF output < 2.0 (nom.)
AF demodulation types BNC, 50 Ω
external reference, external trigger
Connector 0 dBm
10 MHz
Output impedance 2.4 V
0.7 V
Voltage (open circuit)
Chinese, Chinese Traditional, English,
Tracking generator (models .23/.36/.54 only) French, German, Italian, Hungarian,
Japanese, Korean, Portuguese, Russian,
Frequency range Spanish
Port output power 100 kHz ≤ f < 10 MHz SCPI 1997.0
10 MHz ≤ f < 30 GHz 10BASE-T/100BASE-T, RJ-45
VSWR 30 GHz ≤ f < 32 GHz mini B plug, version 2.0
32 GHz ≤ f < 40 GHz
External reference, external trigger 40 GHz ≤ f ≤ 44 GHz WVGA, 800 × 480 pixel
Connector 100 kHz ≤ f < 150 MHz
Mode 150 MHz ≤ f < 8 GHz internal, external headphone supported
External reference 8 GHz ≤ f < 34 GHz type A plug, version 2.0
External trigger threshold 34 GHz ≤ f ≤ 44 GHz 2
USB flash drive, USB version 1.1 or 2.0,
required level size ≤ 32 Gbyte
frequency microSD card, size ≤ 32 Gbyte
low → high transition > 160 instrument settings and traces
high → low transition > 10000 instrument settings and traces
General data
Manual operation
Languages
Remote control number of interfaces
Command set not supplied
LAN interface
USB
Display
Resolution
Audio
Speaker
USB interface
Mass memory
Data storage internal
on USB flash drive or microSD card,
≥ 1 Gbyte
10 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Temperature range operating Version 15.00, May 2022
Climatic loading storage
Altitude battery charging mode –10 °C to +55 °C
Mechanical resistance relative humidity –20 °C to +50 °C
Vibration 0 °C to +40 °C
protection class +25 °C/+55 °C at 95 % relative humidity,
Shock operating with battery in line with EN 60068-2-30
Power supply operating with AC to DC adapter IP51
R&S®HA-Z301 AC power supply 15000 m (49210 ft)
sinusoidal 3000 m (9840 ft)
External DC voltage
Battery random in line with EN 60068-2-6,
MIL-PRF-28800F Class 2
Capacity input specifications in line with EN 60068-2-64,
Voltage MIL-PRF-28800F Class 2
Operating time with new, output specifications 40 g shock spectrum,
fully charged battery operating temperature range in line with MIL-STD-810G, method 516.6
storage temperature range procedure I, MIL-PRF-28800F
Charging time test marks
Life time 100 V to 240 V AC, 50 Hz/60 Hz,
Power consumption R&S®HA-Z306 version E 1.0 A to 0.5 A
R&S®HA-Z306 version F and above 15 V, 2.67 A, max. 40 W
Safety R&S®HA-Z306 version E –30 °C to +60 °C
R&S®HA-Z306 version F and above –40 °C to +85 °C
Test mark R&S®HA-Z306 CE, UL, PSE, TUV
EMC 14.65 V to 15.45 V
model .02 lithium-ion battery
Dimensions model .06 72 Wh
Weight models .13/.26 74.5 Wh
Recommended calibration interval models .23/.36/.44/.54 11.25 V (nom.)
instrument switched off or charge with 10.8 V (nom.)
R&S®HA-Z303 battery charger
instrument switched on 8 h
charging cycles 7 h
6 h
model .02 4.5 h
model .06 3 h
models .13/.26
models .23/.36/.44/.54 5 h
> 75 % or more of its initial capacity after
in line with European EMC Directive 300 charges/discharges
2014/30/EU 8 W (meas.)
10 W (meas.)
W × H × D 12 W (meas.)
16 W (meas.)
models .02/.06/.13/.26 IEC 61010-1:2010/AMD:2016,
models .23/.36/.44/.54 EN 61010-1:2010/A1:2019,
UL. 61010-1 (third edition),
CAN/CSA-C22.2 No. 61010-1:12
VDE, cCSAus, KC
• EN 61326-1
• EN 61326-1 table 2
(immunity, industrial)
• CISPR 11/EN 55011/group 1
Class B (emission)
202 mm × 294 mm × 76 mm
(8.0 in × 11.6 in × 3 in)
2.5 kg (5.5 lb)
3.2 kg (7.1 lb)
1 year
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 11
Version 15.00, May 2022
R&S®FPH-B100 type N RF input connector for model .26
Frequency range model .26 with R&S® FPH-B100 option 5 kHz to 26.5 GHz
Frequency response (R&S®FPH-B31 option is not available in
combination with R&S®FPH-B100 option) < 1.5 dB (nom.)
VSWR temperature range from +20 °C to +30 °C < 1 dB
< 2 dB
5 kHz ≤ f < 10 MHz < 2.5 dB
10 MHz ≤ f < 8 GHz < 2 (nom.)
8 GHz ≤ f < 20 GHz < 1.5 (nom.)
20 GHz ≤ f ≤ 26.5 GHz < 2 (nom.)
100 kHz ≤ f ≤ 100 MHz < 2.7 (nom.)
100 MHz < f ≤ 1 GHz
1 GHz < f ≤ 15.7 GHz
15.7 GHz < f ≤ 26.5 GHz
R&S®FPH-B29 frequency extension from 5 kHz down to 100 Hz
Frequency range models .06/.13/.26/.23/.36/.44/.54, from 5 kHz down to 100 Hz
Frequency response with R&S®FPH-B29 option installed 1 < 3.0 dB (nom.)
temperature range from +20 °C to +30 °C
Displayed average noise level (DANL) –45 dBm (nom.)
100 Hz ≤ f < 5 kHz –100 dBm (nom.)
(only with R&S® FPH-B29 option –45 dBm (nom.)
installed, preamplifier off and –100 dBm (nom.)
attenuator settings ≤ 15 dB)
models .06/.13/.26
preamplifier = off
100 Hz to 100 kHz
100 kHz to 1 MHz
models .23/.36/.44/.54
preamplifier = off
100 Hz to 100 kHz
100 kHz to 1 MHz
R&S®FPH-K7 modulation analysis
Measurement of analog modulation signals (AM, FM)
Center frequency model .02 500 kHz to 2 GHz
500 kHz to 3 GHz
with R&S®FPH-B3 option installed 500 kHz to 4 GHz
with R&S®FPH-B3 and R&S®FPH-B4 500 kHz to 6 GHz
500 kHz to 8 GHz
options installed 500 kHz to 13.6 GHz
500 kHz to 20 GHz
model .06 500 kHz to 26.5 GHz
500 kHz to 31 GHz
with R&S®FPH-B8 option installed 500 kHz to 44 GHz
2 MHz, 1 MHz, 500 kHz, 300 kHz,
models .13/.23 (with tracking generator) 200 kHz, 100 kHz, 50 kHz, 30 kHz,
20 kHz, 10 kHz (nom.)
with R&S®FPH-B20 option installed < ± 5 % (nom.)
models .26/.36 (with tracking generator) carrier power, carrier frequency offset,
AM modulation depth, modulation
with R&S®FPH-B31 option installed frequency, THD, SINAD, SNR
carrier power, carrier frequency offset,
models .44/.54 (with tracking generator) FM deviation, modulation frequency, THD,
SINAD, SNR
Demodulation bandwidth
Bandwidth accuracy
Display AM
FM
12 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Carrier power add 0.2 dB, see level measurement
Carrier power measurement accuracy uncertainty on page 14
0.1 dB
Display resolution
AF (modulation frequency) 5 AM 20 Hz to 100 kHz (nom.)
Range FM 20 Hz to 200 kHz (nom.)
1 Hz
Resolution 1 kHz ≤ AF ≤ 200 kHz ±(1 % of measured value) (nom.)
Measurement uncertainty 20 Hz ≤ AF < 1 kHz ±1 Hz (nom.)
AF filters audio decimation bypass, 1/10, 1/30, 1/100 (nom.)
Lowpass FM demodulation and demodulation off, 50 µs, 75 µs (nom.)
Deemphasis bandwidths 200 kHz and 300 kHz
AM demodulation 6 modulation depth 2 % to 100 % (nom.)
Measurement range frequency deviation ±4 % (nom.)
Modulation depth uncertainty
FM demodulation 7 10 kHz to 400 kHz (nom.),
Measurement range max. 0.4 × demodulation bandwidth
±(0.04 × (AF + deviation)) (nom.)
Deviation uncertainty
Modulation distortion 5, 6, 7
Measurement functions THD, SINAD
–50 dB to 0 dB
Measurement range THD 0 dB to 50 dB
0 dB to 40 dB
SINAD, AM 0.1 dB
1 dB (nom.)
SINAD, FM 20 Hz to 100 kHz (nom.)
Display resolution 10 MHz to 2 GHz
10 MHz to 3 GHz
Measurement uncertainty 10 MHz to 4 GHz
AF frequency range 10 MHz to 6 GHz
10 MHz to 8 GHz
Measurement of digital modulation signals (ASK, FSK) 10 MHz to 13.6 GHz
10 MHz to 20 GHz
Center frequency model .02 10 MHz to 26.5 GHz
10 MHz to 31 GHz
with R&S®FPH-B3 option installed 10 MHz to 44 GHz
400 Hz to 2 MHz,
with R&S®FPH-B3 and R&S®FPH-B4 auto-set corresponding to signal and
demodulation bandwidth requirements
options installed eye diagram, symbols, modulation depth,
modulation error
model .06 carrier power, carrier frequency error,
modulation depth and index,
with R&S®FPH-B8 option installed modulation error
eye diagram, symbols, modulation
models .13/.23 (with tracking generator) deviation, modulation error
carrier power, carrier frequency error,
with R&S®FPH-B20 option installed frequency deviation, modulation error,
magnitude error
models .26/.36 (with tracking generator)
with R&S®FPH-B31 option installed
models .44/.54 (with tracking generator)
Demodulation bandwidth
Display ASK diagram
ASK numerical results
FSK diagram
FSK numerical results
5 Minimum and maximum detectable audio frequency and harmonics depend on the demodulation bandwidth and audio filter settings.
6 Modulation frequency: 1 kHz sine, AM modulation depth: 50 %, carrier level: 0 dBm, center frequency = 499 MHz, reference level: 6 dBm,
demodulation bandwidth = 20 kHz, SNR > 60 dB, audio filter = bypass.
7 Modulation frequency: 1 kHz sine, FM deviation = 75 kHz, carrier level: 0 dBm, center frequency = 499 MHz, reference level: 6 dBm,
demodulation bandwidth = 300 kHz, SNR > 60 dB, audio filter = 1/10, deemphasis = off.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 13
Version 15.00, May 2022
Demodulation parameters transmit filter root raised cosine (RRC)
Modulation and demodulation filters raised cosine (RC)
Gaussian (GAUSS)
Points/symbol unfiltered 8
(Measurement and reference filters are
Filter length internally adapted to signal parameters.)
Demodulation length 4, 8, 16,
internally adapted to signal parameters
internally adapted to signal parameters
20 symbols to max. 1000 symbols
(at 4 points/symbol)
Carrier power add 0.2 dB, see level measurement
Carrier power measurement accuracy uncertainty on page 14
–30 dBm to +20 dBm (nom.)
Carrier power range 0.1 dB
Display resolution
ASK demodulation 9 symbol rate 1 kHz to 100 kHz (nom.)
Measurement range modulation depth 5 % to 95 % (nom.)
±4 % (nom.)
Modulation depth uncertainty 0.1 %
Display resolution
FSK demodulation 10 symbol rate 1 kHz to 100 kHz (nom.)
Measurement range frequency deviation 1 kHz to 400 kHz (nom.)
symbol rate
Accuracy 1 ≤ beta 11 ≤ 20
Display resolution 1 kHz to 20 kHz 1 ≤ beta ≤ 8
> 20 kHz to 50 kHz 1 ≤ beta ≤ 4
> 50 kHz to 100 kHz ±4 % (nom.)
0.1 Hz
R&S®FPH-K19 channel power meter
5 kHz to 2 GHz
Frequency range model .02 5 kHz to 3 GHz
with R&S®FPH-B3 option installed 5 kHz to 4 GHz
Channel bandwidth with R&S®FPH-B3 and R&S®FPH-B4
Amplitude options installed 5 kHz to 6 GHz
Unit 5 kHz to 8 GHz
Limits model .06 5 kHz to 13.6 GHz
Measurement range with R&S®FPH-B8 option installed 5 kHz to 20 GHz
Level measurement uncertainty 5 kHz to 26.5 GHz
Absolute level uncertainty at 100 MHz models .13/.23 (with tracking generator) 5 kHz to 31 GHz
Frequency response (+20 °C to +30 °C) with R&S®FPH-B20 option installed 5 kHz to 44 GHz
100 kHz to 1 GHz
models .26/.36 (with tracking generator) offset, dB relative, zeroing
with R&S®FPH-B31 option installed dBm, W
on/off, upper limit, lower limit, beep on fail
models .44/.54 (with tracking generator) –120 dBm to +30 dBm
+20 °C to +30 °C < 0.3 dB
100 kHz ≤ f < 10 MHz < 1.5 dB (nom.)
10 MHz ≤ f ≤ 4 GHz < 1.25 dB
8 Reference signal is generated with a Gaussian filter, BT = 3.
9 ASK modulation index: 50 %, symbol rate = 100 kHz, Gaussian BT = 1.0, modulation signal: PSBS.
10 FSK modulation deviation: 100 kHz, symbol rate = 100 kHz, Gaussian BT = 1.0, modulation signal: PRBS.
11 Beta is the ratio of frequency deviation to symbol rate.
14 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
R&S®FPH-K29 pulse measurements with power sensor
In combination with one of the R&S®NRP-Z81, R&S®NRP-Z85 or R&S®NRP-Z86 power sensors, the R&S®Spectrum Rider FPH
supports measurements on pulsed signals. The achievable RF performance is documented in the data sheet specifications of the
R&S®NRP-Z81, R&S®NRP-Z85 and R&S®NRP-Z86 power sensors. The list below shows which measurements are supported by the
R&S®FPH-K29.
Measurements R&S®FPH-K29
Pulse power parameters ●
●
Peak power ●
Pulse top power ●
Average power ●
Base power ●
Minimum power ●
Positive overshoot ●
Negative overshoot ●
Pulse timing parameters ●
Pulse duration ●
Pulse period ●
Pulse start/stop time ●
Rise/fall time ●
Duty cycle
R&S®FPH-K43 receiver mode and channel scan measurement application
The list below shows which measurements are supported by the R&S®FPH-K43.
Measurements R&S®FPH-K43
Fixed frequency ●
Frequency scan ●
Channel scan ●
●
User defined channel list ●
EMI precompliance ●
●
CISPR bandwidths
CISPR detectors
The specifications below apply to the R&S®Spectrum Rider FPH. They are based on the data sheet specifications of the
R&S®Spectrum Rider FPH, have not been checked separately and are not verified during instrument calibration.
Frequency range –3 dB bandwidths see basic instrument
Measurement modes –6 dB CISPR bandwidths fixed frequency, frequency scan, channel
scan
Frequency scan stepsize
Scan stepsize 100 Hz to max. frequency
Maximum number of steps 10000
Channel scan user definable
Channel spacing 10000
Maximum number of channels 1 Hz to 3 MHz in 1/3 sequence
200 Hz, 9 kHz, 120 kHz, 1 MHz
Resolution bandwidths max. peak, average, RMS, quasi-peak
see basic instrument
Detectors
Level
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 15
Version 15.00, May 2022
R&S®FPH-K105 EMF measurement application
The list below shows which measurements are supported by the R&S®FPH-K105.
EMF measurements R&S®FPH-K105
Mode Spectrum – channel power (10 MHz)
Sequence of measurements ●
Fixed frequency ●
Set of frequencies ●
Results
Table of field strengths ●
Total field strength ●
Calculation of exposition ●
EMF limit check ●
ISO antenna measurement ●
The setup of the EMF measurement sequence and the detailed result evaluation is done via the R&S®InstrumentView software.
R&S®FPH-K57 advanced gated trigger measurement
Advanced gated trigger measurements are used for analysis of periodic time domain signal measurements and applicable only to the
below mentioned measurement modes.
Measurements R&S®FPH-K57
Occupied bandwidth (OBW) ●
Spectrum emission mask (SEM) ●
Adjacent channel leakage ratio (ACLR) ●
The specifications below apply to the R&S®Spectrum Rider FPH. They are based on the data sheet specifications of the
R&S®Spectrum Rider FPH, have not been checked separately and are not verified during instrument calibration.
Frequency range –3 dB bandwidths see basic instrument
Resolution bandwidths minimum distance, high level to low level 30 kHz to 3 MHz in 1/3 sequence
Video bandwidths 30 kHz to 3 MHz in 1/3 sequence
Detectors see basic instrument
Auto gate detection 10 dB
R&S®HA-Z350 log-periodic OEM antenna
Frequency range 700 MHz to 4 GHz
Gain
Impedance 4 dBi (typ.)
VSWR
Connector type 50 Ω
Dimensions
< 1:2 (nom.)
Weight
Accessories supplied SMA (f)
W × H × D 340 mm × 200 mm × 25 mm
(13.3 in × 7.9 in × 1 in)
270 g (0.6 lb)
hardcase with foam, typical calibration data in 10 MHz steps, pistol grip with mini-tripod
function, one set of SMA toolset
16 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
R&S®FSH-Z14 directional power sensor 12
Frequency range depending on temperature and matching 25 MHz to 1 GHz
Power measurement range (see diagram on page 18) 30 mW to 300 W
VSWR referenced to 50 Ω < 1.06
Power handling capacity 100 W to 1000 W
Insertion loss ≤ 0.06 dB, typ. ≤ 0.04 dB
Directivity > 30 dB
Average power CF: ratio of peak envelope 30 mW to 300 W
Power measurement range power to average power 30 mW to 300 W/CF
CW, FM, PM, FSK, GMSK sine signal 4.0 % of measured value (0.17 dB)
Modulated signals +18 °C to +28 °C, no zero offset 3.2 % of measured value (0.14 dB)
Measurement uncertainty after zeroing ±4 mW
25 MHz to 40 MHz FM, PM, FSK, GMSK 0 % of measured value (0 dB)
40 MHz to 1 GHz AM (80 %) ±3 % of measured value (±0.13 dB)
Zero offset two CW carriers with identical power ±2 % of measured value (±0.09 dB)
Range of typical measurement error EDGE, TETRA ±0.5 % of measured value (±0.02 dB) 13
with modulation 25 MHz to 40 MHz 0.40 %/K (0.017 dB/K)
40 MHz to 1 GHz 0.25 %/K (0.011 dB/K)
Temperature coefficient
Maximum peak envelope power 4 kHz 0.4 W to 300 W
Power measurement range 200 kHz 1 W to 300 W
600 kHz 2 W to 300 W
Video bandwidths same as for average power plus effect of +18 °C to +28 °C
Measurement uncertainty peak hold circuit
Error limits of peak hold circuit for burst duty cycle ≥ 0.1 and repetition rate ≥ 100/s ±(3 % of measured value + 0.05 W),
signals video bandwidth: 4 kHz starting from a burst width of 200 µs
±(7 % of measured value + 0.20 W),
Temperature coefficient video bandwidth: 200 kHz starting from a burst width of 4 µs
±(7 % of measured value + 0.40 W),
video bandwidth: 600 kHz starting from a burst width of 2 µs
plus ±(1.6 % of measured value + 0.15 W)
20/s ≤ repetition rate < 100/s plus ±0.10 W
0.001 ≤ duty cycle < 0.1 0.50 %/K (0.022 dB/K)
25 MHz to 40 MHz 0.35 %/K (0.015 dB/K)
40 MHz to 1 GHz
12 Requires R&S®FSH-Z144 adapter cable.
13 If standard is selected on the R&S®Spectrum Rider FPH.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 17
Version 15.00, May 2022 specifications complied with ≥ 0.4 W 0 dB to 23 dB
> 1.15
Load matching 0.06 W
Matching measurement range
W × H × D 120 mm × 95 mm × 39 mm
Return loss connecting cable (4.72 in × 3.74 in × 1.53 in)
VSWR 1.5 m (59 in)
Minimum forward power 0.65 kg (1.43 lb)
Dimensions and weight
Dimensions
Weight
Error limits for matching measurements Power handling capacity
18 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
R&S®FSH-Z44 directional power sensor 14
Frequency range 200 MHz to 3 GHz 200 MHz to 4 GHz
Power measurement range 3 GHz to 4 GHz 30 mW to 300 W
VSWR referenced to 50 Ω depending on temperature and matching < 1.07
(see diagram on page 20) < 1.12
Power handling capacity 200 MHz to 1.5 GHz 120 W to 1000 W
1.5 GHz to 4 GHz
Insertion loss 200 MHz to 3 GHz ≤ 0.06 dB, typ. ≤ 0.04 dB
3 GHz to 4 GHz ≤ 0.10 dB, typ. ≤ 0.07 dB
Directivity > 30 dB
> 26 dB
Average power CF: ratio of peak envelope power to average power
Power measurement range
CW, FM, PM, FSK, GMSK 30 mW to 300 W
Measurement uncertainty
Zero offset LTE, 3GPP WCDMA, cdmaOne, 30 mW to 120 W
Range of typical measurement error
with modulation CDMA2000®, DAB, DVB-T
Temperature coefficient other modulated signals 30 mW to 300 W/CF
sine signal, +18 °C to +28 °C, no zero offset
200 MHz to 300 MHz 4.0 % of measured value (0.17 dB)
300 MHz to 4 GHz 3.2 % of measured value (0.14 dB)
after zeroing ±4 mW
FM, PM, FSK, GMSK 0 % of measured value (0 dB)
AM (80 %) ±3 % of measured value (±0.13 dB)
two CW carriers with identical power ±2 % of measured value (±0.09 dB)
π/4-DQPSK ±2 % of measured value (±0.09 dB)
EDGE ±0.5 % of measured value (±0.02 dB) 15
cdmaOne, DAB ±1 % of measured value (±0.04 dB) 11
3GPP WCDMA, CDMA2000® ±2 % of measured value (±0.09 dB) 11
DVB-T ±2 % of measured value (±0.09 dB) 11
200 MHz to 300 MHz 0.40 %/K (0.017 dB/K)
300 MHz to 4 GHz 0.25 %/K (0.011 dB/K)
Maximum peak envelope power 4 W to 300 W
Power measurement range
4 kHz 0.4 W to 300 W
DAB, DVB-T, cdmaOne, CDMA2000®,
3GPP WCDMA 200 kHz 1 W to 300 W
Other signals at video bandwidth
4 MHz 2 W to 300 W
Measurement uncertainty
Error limits of peak hold circuit for burst +18 °C to +28 °C same as for average power plus effect of
signals
peak hold circuit
Range of typical measurement error of
peak hold circuit duty cycle ≥ 0.1 and repetition rate ≥ 100/s
Temperature coefficient
video bandwidth 4 kHz ±(3 % of measured value + 0.05 W)
starting from a burst width of 100 µs
video bandwidth 200 kHz ±(3 % of measured value + 0.20 W)
starting from a burst width of 4 µs
video bandwidth 4 MHz ±(7 % of measured value + 0.40 W)
starting from a burst width of 1 µs
20/s ≤ repetition rate < 100/s plus ±(1.6 % of measured value + 0.15 W)
0.001 ≤ duty cycle < 0.1 plus ±0.10 W
burst width ≥ 0.5 µs plus ±5 % of measured value
burst width ≥ 0.2 µs plus ±10 % of measured value
video bandwidth 4 MHz and standard selected on the R&S®FPH4/8/13/20
cdmaOne, DAB ±(5 % of measured value + 0.4 W)
DVB-T, CDMA2000®, 3GPP WCDMA ±(15 % of measured value + 0.4 W)
200 MHz to 300 MHz 0.50 %/K (0.022 dB/K)
300 MHz to 4 GHz 0.35 %/K (0.015 dB/K)
14 Requires R&S®FSH-Z144 adapter cable.
15 If standard is selected on the R&S®Spectrum Rider FPH.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 19
Version 15.00, May 2022 200 MHz to 3 GHz 0 dB to +23 dB
3 GHz to 4 GHz 0 dB to +20 dB
Load matching 200 MHz to 3 GHz > 1.15
Matching measurement range 3 GHz to 4 GHz > 1.22
specifications complied with ≥ 0.2 W 0.03 W
Return loss
VSWR W × H × D 120 mm × 95 mm × 39 mm
VSWR connecting cable (4.72 in × 3.74 in × 1.53 in)
1.5 m (59 in)
Minimum forward power 0.65 kg (1.43 lb)
Dimensions and weight
Dimensions
Weight
Error limits for matching measurements Power handling capacity
20 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Equivalence of specifications for different R&S®Spectrum Rider FPH
part numbers
• The specifications for part number 1321.1111.02 are equivalent to part number 1321.1111.52 and 1321.1111P01.
• The specifications for part number 1321.1111.06 are equivalent to part number 1321.1111.56 and 1321.1111P04.
• The specifications for part number 1321.1111.13 are equivalent to part number 1321.1111.63 and 1321.1111P06.
• The specifications for part number 1321.1111.26 are equivalent to part number 1321.1111.76 and 1321.1111P08.
• The specifications for part number 1321.1711.23 are equivalent to part number 1321.1711P02.
• The specifications for part number 1321.1711.36 are equivalent to part number 1321.1711P04.
• The specifications for part number 1321.1711.44 are equivalent to part number 1321.1711P06.
• The specifications for part number 1321.1711.36 are equivalent to part number 1321.1711P07.
Ordering information
Designation Type Order No.
Handheld spectrum analyzer, 5 kHz to 2 GHz R&S®Spectrum Rider FPH 1321.1111.02
Handheld spectrum analyzer, 5 kHz to 6 GHz R&S®Spectrum Rider FPH 1321.1111.06
Handheld spectrum analyzer, 5 kHz to 13.6 GHz R&S®Spectrum Rider FPH 1321.1111.13
Handheld spectrum analyzer, 5 kHz to 26.5 GHz R&S®Spectrum Rider FPH 1321.1111.26
Handheld spectrum analyzer, 5 kHz to 44 GHz R&S®Spectrum Rider FPH 1321.1711.44
Handheld spectrum analyzer, 5 kHz to 13.6 GHz, R&S®Spectrum Rider FPH 1321.1711.23
with tracking generator
Handheld spectrum analyzer, 5 kHz to 26.5 GHz, R&S®Spectrum Rider FPH 1321.1711.36
with tracking generator
Handheld spectrum analyzer, 5 kHz to 44 GHz, R&S®Spectrum Rider FPH 1321.1711.54
with tracking generator
Accessories supplied
Lithium-ion battery pack, USB cable, AC power supply with country specific adapters for EU, GB, US, AUS, CH, documentation,
quick start guide, side strap
Options Type Order No.
R&S®FPH-B3 1321.0667.02
Designation R&S®FPH-B4 1321.0673.02
Spectrum analyzer frequency upgrade, 2 GHz to 3 GHz 16
Spectrum analyzer frequency upgrade, 3 GHz to 4 GHz 16 R&S®FPH-B8 1321.0767.02
(requires R&S®FPH-B3) R&S®FPH-B20 1321.0773.02
Spectrum analyzer frequency upgrade, 6 GHz to 8 GHz 17 R&S®FPH-B31 1321.0780.02
Spectrum analyzer frequency upgrade, 13.6 GHz to 20 GHz 18 R&S®FPH-B22 1321.0680.02
Spectrum analyzer frequency upgrade, 26.5 GHz to 31 GHz 19, 20 R&S®FPH-B23 1321.0867.02
Spectrum analyzer preamplifier, 5 kHz to 4 GHz 16 R&S®FPH-B24 1321.0850.02
Spectrum analyzer preamplifier, 5 kHz to 8 GHz 17 R&S®FPH-B25 1321.0873.02
Spectrum analyzer preamplifier, 5 kHz to 20 GHz 18 R&S®FPH-B26 1334.6600.02
Spectrum analyzer preamplifier, 5 kHz to 31 GHz 19 R&S®FPH-B100 1321.0596.02
Spectrum analyzer preamplifier, 5 kHz to 44 GHz 21 R&S®FPH-B29 1334.8532.02
Type N RF input connector, for model .26 (factory installed) 20
Spectrum analyzer 100 Hz frequency extension, R&S®FPH-K7 1321.0696.02
from 5 kHz down to 100 Hz 22 R&S®FPH-K9 1321.0709.02
Analog modulation analysis AM, FM, ASK, FSK R&S®FPH-K15 1321.0715.02
Power sensor support R&S®FPH-K16 1321.0615.02
Interference analysis R&S®FPH-K19 1321.0721.02
Signal strength mapping R&S®FPH-K29 1321.0738.02
Channel power meter R&S®FPH-K43 1321.0621.02
Pulse measurements, with power sensor R&S®FPH-K57 1321.1586.02
Receiver mode and channel scanner R&S®FPH-K105 1334.6946.02
Advanced gated trigger measurements
EMF measurement
16 Applicable only to base unit with order no. 1321.1111.02.
17 Applicable only to base unit with order no. 1321.1111.06.
18 Applicable only to base unit with order no. 1321.1111.13 and 1321.1711.23.
19 Applicable only to base unit with order no. 1321.1111.26 and 1321.1711.36.
20 R&S®FPH-B31 option is not available in combination with R&S®FPH-B100 option.
21 Applicable only to base unit with order no. 1321.1711.44 and 1321.1711.54.
22 For serial number ≥ 103100. Not applicable to R&S®Spectrum Rider FPH model .02.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 21
Version 15.00, May 2022
Extras Type Order No.
R&S®HA-Z303 1321.1328.02
Designation R&S®HA-Z306 1321.1334.02
Battery charger for R&S®HA-Z306 23 R&S®HA-Z301 1321.1386.02
Lithium-ion battery pack, 6.4 Ah R&S®HA-Z302 1321.1340.02
Spare power supply, incl. mains plug for EU, GB, US, AUS, CH R&S®HA-Z322 1321.1370.02
Car adapter R&S®HA-Z322 1321.1370.03
Carrying holster R&S®HA-Z220 1309.6175.00
Rainproof carrying holster R&S®HA-Z321 1321.1357.02
Soft carrying bag R&S®RTH-Z4 1326.2774.02
Hardcase R&S®FSH-Z36 1145.5838.02
Hard shell protective carrying case R&S®HA-Z211 1309.6169.00
Headphones R&S®HA-Z210 1309.6152.00
Spare USB cable
Spare Ethernet cable Order No.
1328.6825.02
Antennas and antenna accessories Type 1328.6283.02
R&S®HA-Z1900 4500.7470.00
Designation R&S®HA-Z900 3626.2757.02
Yagi antenna, 1710 MHz to 1990 MHz R&S®CS-ZANT
Yagi antenna, 824 MHz to 960 MHz R&S®HA-Z901 1328.6883.02
Telescopic antenna
RF cable, length: 1 m, DC to 6 GHz, R&S®HA-Z902 4104.6000.04
type N (m) to type N (m) connectors 4104.7770.04
Carrying bag, R&S®HE400BC 4104.6000.02
for R&S®HA-Z900 or R&S®HA-Z1900 Yagi antenna R&S®HE400-KB 4104.6000.03
Basic handheld directional antenna (antenna handle) R&S®HE400 4104.7770.02
RF cable, for R&S®HE400BC R&S®HE400MW
Handheld directional antenna (antenna handle) R&S®HE400-K 4104.8002.02
Microwave handheld directional antenna (antenna handle) 4104.8202.02
Cable set, for R&S®HE400 and R&S®HE400MW R&S®HE400HF 4104.6900.02
(requires R&S®HE300USB) R&S®HE400VHF 4104.8402.02
HF antenna module, 8.3 kHz to 30 MHz R&S®HE400UWB 4104.7306.02
VHF antenna module, 20 MHz to 200 MHz R&S®HE400LP 4104.7606.02
UWB antenna module, 30 MHz to 6 GHz R&S®HE400CEL 4104.8602.02
Log-periodic antenna module, 450 MHz to 8 GHz R&S®HE400SCB
Cellular antenna module, 700 MHz to 2500 MHz R&S®HE400SHF 4080.9440.02
S band and C band antenna module, 1.7 GHz to 6 GHz 1321.1405.02
SHF antenna module, 5 GHz to 20 GHz (with R&S®HE400BC R&S®HE300USB 4115.6006.02
and R&S®HE400MW antenna handle) R&S®HA-Z350 4115.7660.02
USB adapter, for R&S®HE400 handheld directional antenna R&S®HE800-PA 1309.6600.00
Log-periodic OEM antenna, 700 MHz to 4 GHz R&S®HE800Z1
Handheld directional antenna, with preamplifier R&S®FSH-Z320 1309.6617.00
Transport case, for R&S®HE800-PA
RF cable, length: 1 m, DC to 8 GHz, armored, R&S®FSH-Z321 1321.1392.02
type N (m) to type N (f) connectors 1158.9295.05
RF cable, length: 3 m, DC to 8 GHz, armored, R&S®HA-Z340 1074.5719.02
type N (m) to type N (f) connectors R&S®TS-EMF 1074.5702.02
GPS receiver, for R&S®Spectrum Rider FPH R&S®TSEMF-B1 1074.5690.02
Portable EMF measurement system, hard case R&S®TSEMF-B2 1158.9250.02
Isotropic antenna, 30 MHz to 3 GHz for R&S®TS-EMF R&S®TSEMF-B3 0358.5414.02
Isotropic antenna, 700 MHz to 6 GHz for R&S®TS-EMF R&S®TSEMF-CV 0358.5714.02
Isotropic antenna, 9 kHz to 200 MHz for R&S®TS-EMF R&S®RAM 1300.7740.02
Converter cable R&S®RAZ 0118.2812.00
Matching pad, 50/75 Ω, L section R&S®FSH-Z38 0092.6581.00
Matching pad, 50/75 Ω, series resistor 25 Ω 4012.5837.00
Matching pad, 50/75 Ω, L section, type N to BNC 3692.7660.00
Adapter type N (m) to BNC (f) 3530.6646.00
Adapter type N (m) to type N (m) 3530.6630.00
Adapter type N (m) to SMA (f) 4048.9790.00
Adapter type N (f) to SMA (f)
Adapter type N (m) to 7/16 (f)
Adapter type N (m) to 7/16 (m)
Adapter type N (m) to FME (f)
23 The battery charger is dedicated for charging an additional battery outside the instrument. The battery can be charged via the instrument as well.
22 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Designation Type Version 15.00, May 2022
Adapter BNC (m) to banana (f)
Attenuator, 50 W, 20 dB, 50 Ω, DC to 6 GHz, R&S®RDL50 Order No.
type N (f) to type N (m) R&S®RBU100 0017.6742.00
Attenuator, 100 W, 20 dB, 50 Ω, DC to 2 GHz, R&S®RBU100 1035.1700.52
type N (f) to type N (m) R&S®HZ-15 1073.8495.20
Attenuator, 100 W, 30 dB, 50 Ω, DC to 2 GHz, R&S®HZ-17
type N (f) to type N (m) R&S®HZ-16 1073.8495.30
Compact probe set, for E and H near-field measurements, R&S®AC004R1
30 MHz to 3 GHz R&S®AC004L1 1147.2736.02
Near-field probe set, H field R&S®AC004R2
Preamplifier (3 GHz, 20 dB), power adapter (100 V to 230 V), R&S®AC004L2 1339.4141.02
for R&S®HZ-15 R&S®HF907OM 1147.2720.02
Omnidirectional antenna, for circular right-hand polarization, R&S®FH-SG-40
18 GHz to 26.5 GHz R&S®HA-Z370 0749.3000.03
Omnidirectional antenna, for circular left-hand polarization, R&S®KM011Z8
18 GHz to 26.5 GHz R&S®HZ-1 4078.4000.02
Omnidirectional antenna, for circular right-hand polarization, R&S®ZV-Z93 0749.3251.03
26.5 GHz to 40 GHz R&S®ZV-Z93
Omnidirectional antenna, for circular left-hand polarization, R&S®ZV-Z193 4078.5006.02
26.5 GHz to 40 GHz R&S®ZV-Z193
Broadband omnidirectional antenna, 800 MHz to 26.5 GHz R&S®ZV-Z193 4070.3279.02
Standard gain horn antenna, 26 GHz to 40 GHz, R&S®ZV-Z95 3629.2393.02
mid band gain 20 dB, WR 28 R&S®ZV-Z95
Standard gain horn antenna adapter R&S®ZV-Z195 1334.8432.02
Mast and tripod adapter R&S®ZV-Z195 4090.4006.02
Wooden tripod 0837.2310.02
Test port cable, 0 Hz to 26.5 GHz, 1301.7595.25
3.5 mm (f) to 3.5 mm (m), length: 635 mm
Test port cable, 0 Hz to 26.5 GHz, 1301.7595.38
3.5 mm (f) to 3.5 mm (m), length: 965 mm 1306.4520.24
Test port cable, 0 Hz to 26.5 GHz, 1306.4520.36
3.5 mm (f) to 3.5 mm (m), length: 610 mm 1306.4520.60
Test port cable, 0 Hz to 26.5 GHz, 1301.7608.25
3.5 mm (f) to 3.5 mm (m), length: 914 mm 1301.7608.38
Test port cable, 0 Hz to 26.5 GHz, 1306.4536.24
3.5 mm (f) to 3.5 mm (m), length: 1524 mm 1306.4536.36
Test port cable, 0 Hz to 40 GHz,
2.92 mm (f) to 2.92 mm (m), length: 635 mm
Test port cable, 0 Hz to 40 GHz,
2.92 mm (f) to 2.92 mm (m), length: 965 mm
Test port cable, 0 Hz to 40 GHz,
2.92 mm (f) to 2.92 mm (m), length: 610 mm
Test port cable, 0 Hz to 40 GHz,
2.92 mm (f) to 2.92 mm (m), length: 914 mm
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 23
Version 15.00, May 2022
Supported R&S®NRP-Zxx power sensors 24
Designation Type Order No.
Directional power sensor, 25 MHz to 1 GHz R&S®FSH-Z14 1444.0029.02
Directional power sensor, 200 MHz to 4 GHz R&S®FSH-Z44 1444.0035.02
Universal power sensor, 10 MHz to 8 GHz, 100 mW, two-path R&S®NRP-Z211 1417.0409.02
Universal power sensor, 10 MHz to 18 GHz, 100 mW, two-path R&S®NRP-Z221 1417.0309.02
Wideband power sensor, 50 MHz to 18 GHz, 100 mW R&S®NRP-Z81 1137.9009.02
Wideband power sensor, 50 MHz to 40 GHz, 100 mW (2.92 mm) R&S®NRP-Z85 1411.7501.02
Wideband power sensor, 50 MHz to 40 GHz, 100 mW (2.40 mm) R&S®NRP-Z86 1417.0109.40
Wideband power sensor, 50 MHz to 44 GHz, 100 mW (2.40 mm) R&S®NRP-Z86 1417.0109.44
Three-path diode power sensor, 100 pW to 200 mW, R&S®NRP8S 1419.0006.02
10 MHz to 8 GHz
Three-path diode power sensor, 100 pW to 200 mW, R&S®NRP18S 1419.0029.02
10 MHz to 18 GHz
Three-path diode power sensor, 100 pW to 200 mW, R&S®NRP33S 1419.0064.02
10 MHz to 33 GHz
Three-path diode power sensor, 100 pW to 200 mW, R&S®NRP40S 1419.0041.02
10 MHz to 40 GHz
Three-path diode power sensor, 100 pW to 200 mW, R&S®NRP50S 1419.0087.02
10 MHz to 50 GHz
Thermal power sensor, 300 nW to 100 mW, DC to 18 GHz R&S®NRP18T 1424.6115.02
Thermal power sensor, 300 nW to 100 mW, DC to 33 GHz R&S®NRP33T 1424.6138.02
Thermal power sensor, 300 nW to 100 mW, DC to 40 GHz R&S®NRP40T 1424.6150.02
Thermal power sensor, 300 nW to 100 mW, DC to 50 GHz R&S®NRP50T 1424.6173.02
Thermal power sensor, 300 nW to 100 mW, DC to 67 GHz R&S®NRP67T 1424.6196.02
Thermal power sensor, 300 nW to 100 mW, DC to 110 GHz R&S®NRP110T 1424.6215.02
Average power sensor, 100 pW to 200 mW, 8 kHz to 6 GHz R&S®NRP6A 1424.6796.02
Average power sensor, 100 pW to 200 mW, 8 kHz to 18 GHz R&S®NRP18A 1424.6815.02
R&S®NRP-Zxx power sensors require the following adapter cable for operation on the R&S®Spectrum Rider FPH
USB adapter cable, for R&S®FSH-Z14/R&S®FSH-Z44 R&S®FSH-Z144 1145.5909.02
power sensors
USB adapter cable (passive), length: 2 m, R&S®NRP-Z4 1146.8001.02
to connect R&S®NRP-Zxx S/SN power sensors to the
R&S®Spectrum Rider FPH
R&S®NRP power sensors require the following adapter cable for operation on the R&S®Spectrum Rider FPH
USB interface cable, length: 1.5 m, R&S®NRP-ZKU 1419.0658.03
to connect R&S®NRP sensors to the R&S®Spectrum Rider FPH
Optical power sensors and accessories
Designation Type Order No.
OEM USB, optical power meter (Germanium) R&S®HA-Z360 1334.5162.00
OEM USB, optical power meter (filtered InGaAs) R&S®HA-Z361 1334.5179.00
SC adapter, for optical power meter R&S®HA-Z362 1334.5185.00
LC adapter, for optical power meter R&S®HA-Z363 1334.5191.00
2.5 mm universal adapter, for optical power meter R&S®HA-Z364 1334.5204.00
1.25 mm universal adapter, for optical power meter R&S®HA-Z365 1334.5210.00
Patch cord SC-LC SM, SX, length: 1 m R&S®HA-Z366 1334.5227.00
Patch cord SC-SC SM, SX, length: 1 m R&S®HA-Z367 1334.5233.00
24 For average power measurements only.
24 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Service options R&S®WE1 3 years
R&S®WE2 1 year
Warranty R&S®CW1
Base unit R&S®CW2 Please contact your local
All other items 25 R&S®AW1 Rohde & Schwarz sales office.
Service options
R&S®AW2
Extended warranty, one year
Extended warranty, two years
Extended warranty with calibration coverage, one year
Extended warranty with calibration coverage, two years
Extended warranty with accredited calibration coverage,
one year
Extended warranty with accredited calibration coverage,
two years
Extended warranty with a term of one and two years (WE1 and WE2)
Repairs carried out during the contract term are free of charge 26. Necessary calibration and adjustments carried out during repairs are
also covered.
Extended warranty with calibration coverage (CW1 and CW2)
Enhance your extended warranty by adding calibration coverage at a package price. This package ensures that your
Rohde & Schwarz product is regularly calibrated, inspected and maintained during the term of the contract. It includes all repairs 26 and
calibration at the recommended intervals as well as any calibration carried out during repairs or option upgrades.
Extended warranty with accredited calibration (AW1 and AW2)
Enhance your extended warranty by adding accredited calibration coverage at a package price. This package ensures that your
Rohde & Schwarz product is regularly calibrated under accreditation, inspected and maintained during the term of the contract. It
includes all repairs 26 and accredited calibration at the recommended intervals as well as any accredited calibration carried out during
repairs or option upgrades.
CDMA2000® is a registered trademark of the Telecommunications Industry Association (TIA-USA).
25 For options that are installed, the remaining base unit warranty applies if longer than 1 year. Exception: all batteries have a 1 year warranty.
26 Excluding defects caused by incorrect operation or handling and force majeure. Wear-and-tear parts are not included.
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 25
Version 15.00, May 2022
26 Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Version 15.00, May 2022
Rohde & Schwarz R&S®Spectrum Rider FPH Handheld Spectrum Analyzer 27
Service that adds value Rohde & Schwarz
The Rohde & Schwarz technology group is among the trail-
► Worldwide blazers when it comes to paving the way for a safer and
► Local and personalized connected world with its leading solutions in test & measure-
► Customized and flexible ment, technology systems and n etworks & cybersecurity.
► Uncompromising quality Founded more than 85 years ago, the group is a reliable
► Long-term dependability partner for industry and government customers around
the globe. The independent company is headquartered in
Munich, Germany and has an extensive sales and service
network with locations in more than 70 countries.
www.rohde-schwarz.com
Sustainable product design
► Environmental compatibility and eco-footprint
► Energy efficiency and low emissions
► Longevity and optimized total cost of ownership
Certified Quality Management Certified Environmental Management
ISO 9001 ISO 14001
Rohde & Schwarz training
www.training.rohde-schwarz.com
Rohde & Schwarz customer support
www.rohde-schwarz.com/support
3607214922
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG 3607.2149.22 15.00 PDP/PDW 1 en
Trade names are trademarks of the owners
PD 3607.2149.22 | Version 15.00 | May 2022 (ja)
R&S®Spectrum Rider FPH Handheld Spectrum Analyzer
Data without tolerance limits is not binding | Subject to change
© 2015 - 2022 Rohde & Schwarz GmbH & Co. KG | 81671 Munich, Germany