Rad-5®
A handheld pulse oximeter with Masimo SET® Measure-through
Motion and Low Perfusion™ pulse oximetry
> In a study published in Journal of Clinical Anesthesiology comparing three pulse oximetry technologies, Masimo SET® demonstrated
the highest sensitivity and specificity in identifying desaturation events during conditions of motion and low perfusion1
> Lightweight, handheld device with user-configurable power up default settings
> Sleep Mode allows clinicians to disable audio tones and dim LEDs
> Up to 72 hours of trending memory
> Perfusion Index (Pi) is an assessment of the pulsatile strength at a specific monitoring site (e.g. the
hand, finger, or foot), and as such Pi is an indirect and noninvasive measure of peripheral perfusion
> Signal I.Q.® (SIQ) provides an assessment of the confidence in the displayed SpO2 value
> FastSat® tracks rapid changes in arterial O2
> SmartTone beeps in sync with pulse, even under patient motion conditions
> Sensitivity options: APOD®, Normal, and MAX
> Audible and visual alarms for High/Low Saturation,Pulse Rate, Sensor Off, and Low Battery
Features
FastSat tracks rapid changes in arterial O2. The Alarm Status Indicator flashes when an © 2017 Masimo. All rights reserved.
alarm condition is present.
FastSat TM FastSat TM
FastSat TM
>40 >5 >40 >5 >40 >5
35 Signal I.Q. (%SSIQpO)2 provide32 s an assessm3350ent of 3 Per3f5usion Index (Pi) is an assessment of the 3
30 30
%SpO2 2 %SpO2 2
pul2s5atile strength at a specific monitoring
25 the confidence in t1h.75e displayed SpO2 v2a5lue. A 1.75 20 1.75
20 vertical LED bar ris11.e2.55s and falls with the1250pulse, 1.5 site15(e.g. the hand, finger, or foot), and as such 1.5
15
1.25 1.25
12 where the height of th.15e bar indicat1e92s the 1 Pi is12 an indirect and noninvasive measure of 1
9 9
.5 .5
6 quality of the signal (lef<.2t.51graphic). 6 .25 per<i63pheral perfusion. The LED bar is highest .25
<3 <3
<.1 <.1
SIQTM BPM PI SIQTM BPM PI and green when the qu SIQTMality of the perfused BPM PI
MAX MAX site is best; when Pi is poor the LED bar is low MAX
and turns red (right graphic).
Protective boots are available in your choice of seven different colors.
Rad-5 Specifications
MEASUREMENT RANGE PHYSICAL CHARACTERISTICS
SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–100% Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2” x 3.0” x 1.4” (15.8 cm x 7.6 cm x 3.6 cm)
Pulse Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–240 bpm Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 oz (0.32 kg)
Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.02–20%
TRENDING
ACCURACY (ARMS)2
Provides up to 72 hours of trending at 2 second resolution.
Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70–100% Output to PC running Masimo TrendCom™ Utility
No Motion Adults/Pediatrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 %
No Motion Neonates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 % MODES
Motion Adults/Pediatrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 %
Motion Neonates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 % Averaging Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 4, 8, 10, 12, 14, or 16 seconds
Low Perfusion Adults/Pediatrics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 % Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . APOD, Normal, and Maximum
Low Perfusion Neonates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 %
ALARMS
Pulse Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25–240 bpm
No Motion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 bpm Audible and visual alarms for high and low saturation and pulse rate (SpO2
Motion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 bpm range 1% - 100%, pulse rate range 25 - 240 bpm)
Low Perfusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 bpm Sensor condition, system failure, and low battery alarms
High Priority. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 799 Hz tone, 5 pulse burst,
RESOLUTION
pulse spacing: 0.250s, 0.250s, 0.500s, 0.250s, repeat time: 10s
Saturation (%SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1% Low Priority . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 432 Hz tone, 3 pulses, repeat time: 5s
Pulse Rate (bpm).. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 bmp Alarm Volume . . . . . . . . . . . . . . . . . . . . . . High Priority: 75 dB (max), Low Priority: 75 dB (max)
BATTERIES DISPLAY/INDICATORS
Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 AA Alkaline Data Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . % SpO2, pulse rate, perfusion index,
Capacity.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . over 30 hours FastSat, alarm status, alarm silenced status,
Signal IQ/pleth bar, battery status, MAX
ENVIRONMENTAL
Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LED
Operating Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32°F to 122°F (0°C to 50°C)
Storage Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40°F to 158°F (-40°C to 70°C) COMPLIANCE
Operating Humidity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5% to 95%, non-condensing
Operating Altitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mbar to 1060 mbar pressure EMC Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IEC 60601-1 2, Class B
Equipment Classification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IEC 60601-1-1 / UL 60601-1
-1000 ft to 18,000 ft (-304 m to 5,486 m) Type of Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Internally powered (on battery power)
Degree of Protection-Patient Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Type BF-Applied Part
Rad-5 Mode of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
1 Shah et al. J Clin Anesth. 2012;24(5):385-91. 2 ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements.
Approximately two-thirds of the device measurements fell within ± ARMS of the reference measurements in a controlled study.
Caution: Federal (USA) law restricts this device to sale by or on the order Masimo U.S. Masimo International 7322/PLM-10658A-0417
of a physician. See instructions for use for full prescribing information,
including indications, contraindications, warnings, and precautions. Tel: 1 877 4 Masimo Tel: +41 32 720 1111
info-america@masimo.com info-international@masimo.com